16 Special Functions
16.6 GC Configuration
240
GC-2014
AIR
Selection: Yes/No, Default: Yes
Indicate which detector APC zones should be included in the check.
The displayed items depend on the detector type.
AUX APC
APC1APC15 PRESS
Selection: Yes/No, Default: Yes
This item is available only when optional APC units are installed.
AUX AMC
AMC1AMC10 FLOW
Selection: Yes/No, Default: Yes
This item is available only when optional AMC units are installed.
DETECTOR
CONTROLLER
Selection: Yes/No, Default: Yes
This item is available only when the FID is installed.
DRIFT
Selection: Yes/No, Default: Yes
This item compares the baseline drift to the drift limit for 10 minutes.
Once the GC becomes ready, the GC re-evaluates of the ready status 10 minutes later.
WAIT SIGNAL
Selection: Yes/No, Default: Yes
The wait signal applies to all detectors configured in analytical lines.
User-specified names are used in this screen.
16.6.6.3 PF menu
PF menu Description
Reference
section
Drift Set the baseline drift limit.
――
NOTE

Table of Contents

Other manuals for Shimadzu GC-2014

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Shimadzu GC-2014 and is the answer not in the manual?

Shimadzu GC-2014 Specifications

General IconGeneral
TypeGas Chromatograph
DisplayLCD
Maximum Heating Rate120°C/min
Data AcquisitionDigital
Maximum Number of Detectors4
WeightApprox. 45 kg
Carrier GasHelium, Nitrogen, Hydrogen
Detector OptionsFID, TCD, ECD, FPD
Injection PortSplit/Splitless, On-column
Gas ControlElectronic Flow Control (EFC)
Temperature Programming Rate0.1 to 120 °C/min
Power Requirements230 V AC, 50/60 Hz
Detector Temperature RangeUp to 400°C (depending on detector type)