Shimadzu GC-2014 - Maintenance Parts and Consumables; 6.1 Consumable Parts

Shimadzu GC-2014
318 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
4 Installing Capillary Columns and Setting Analytical Flow Lines
4.3 Capillary Analysis Using the Dual FID (When a makeup gas flow controller is used)
53
GC-2014
Setting analytical flow lines
Setting analytical flow lines is described below.
For procedures to install an adopter and columns and input analytical conditions, refer to the
operation manual.
1. Set a line.
(1) Press the [SET] key and then press the [PF2] key (Line Config).
(2) Set the SPL and dual FID (DFID) on a same line.
LINE1 is specified in the example below. Lines 1 through 4 can be selected.
2. Set the dual FID.
(1) Press the [DET] key.
(2) Set “Control mode” as described below.
Set to “Single L” when the column is connected to the left (L) side of the dual FID and “Single
R” when it is connected to the right (R) side.