Page #206 background imageLoading...
Page #206 background image
Display the object history
The object history shows the audit history for the object and for all objects contained
in the selected object. For example, when you create an assay, a copy of the instrument
protocol and the primary analysis protocol (and therefore dye set, and size standard)
are included in the assay object. The objects contained within an object have audit
histories distinct from the audit history of the objects stored in the Library.
To display the history for an object:
Select the object, then click Show Object History.
System configuration history
Table 1 Audit – System configuration history
Record Type Action Corresponds to
Security settings Update
Enable security
Disable security
Modify security policies:
Session timeout settings
Account settings Update
Modify user name settings
Modify password settings
Modify security policies:
Password expiration
Account suspension
Audit reason for
change
Update Modify reason for change
Create Create reason for change
Delete Delete reason for change
Audit settings Update
Enable auditing
Disable auditing
Audit type Update Modify audit settings
E-Signature
function
Update
Modify the authorities for a “prompt
before” function
Modify the Enable state of either a
“check after” or “prompt before”
function
E-Signature settings Update
Enable e-signature
Disable e-signature
E-Signature type Update
Modify e-signature settings
Modify the enable state of an E-
Signature Type
Chapter 8 Use Security, Audit, and E-Sig functions (SAE Module)
Administrators
8
206
3500/3500xL Genetic Analyzer User Guide—Data Collection Software v3.1

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Applied Biosystems 3500 and is the answer not in the manual?

Applied Biosystems 3500 Specifications

General IconGeneral
Number of capillaries8
Capillary Length50 cm
LaserSolid-state laser
Instrument TypeCapillary electrophoresis system
TechnologyCapillary Electrophoresis (CE)
Weight95 kg
Power Requirements100–240 V