EasyManua.ls Logo

Applied Biosystems 3500 - Flush the Water Trap (Pump Trap)

Applied Biosystems 3500
326 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
The Wash Pump and Channels wizard takes >40 minutes to complete.
1.
In the Dashboard, click W
izards.
2.
In the Maintenance wizards screen, click Wash Pump and Channels.
3.
Follow the prompts in the Wash wizard window.
Flush the water trap monthly to prolong the life of the pump and to remove diluted
polymer from the pump.
Flush with distilled or deionized w
ater and ensure that the water ows into the
overow container. Dispose of the excess water (inside the overow container). See
“Chemical safety“ on page 308.
Note: Leave the trap lled with either distilled or deionized water.
1.
Fill the supplied 20 mL, all-plastic Luer lock syringe (in the Polymer Delivery
Pump Cleaning Kit , Cat. No. 4414007 ) with distilled or deionized water. Expel
any bubbles from the syringe.
IMPORTANT! Do not use a syringe smaller than 20 mL. Doing so may generate
excessiv
e pressure within the trap.
Flush the water
trap (pump trap)
Chapter 9 Maintain the Instrument
Maintain the pump
9
238
3500/3500xL Genetic Analyzer User Guide—Data Collection Software v3.1

Table of Contents

Related product manuals