Clean the instrument
IMPORTANT! Wear appropriate protection, including gloves, laboratory goggles, and
coat whenever you work with the uids used on this instrument, or parts that may
come into contact with these uids.
IMPORTANT! Use only the cleaning agents listed in this guide. Use of cleaning
agents other than those listed in this guide may damage the instrument.
1.
Ensure the ov
en is closed.
2.
Press the Tray buon on the front of the instrument to move the autosampler to
the forward position.
3.
Wipe o any liquid on or around the autosampler using a lint-free tissue.
4.
Clean o any polymer build-up crystals on the instrument, including the
capillary tips, with deionized water and lint-free tissue.
5.
Clean the array plug with deionized water and lint-free tissue.
6.
Clean out the drip trays with deionized water, or ethanol, and lint-free tissue.
Note: The drip tray can be removed.
Install buffers
IMPORTANT! W
ear appropriate protection, including gloves, laboratory goggles, and
coat whenever you work with the uids used on this instrument, or parts that may
come into contact with these uids.
IMPORTANT! Use only the parts listed in Appendix D, “Catalog numbers“.
1.
Check the expiration date on the label to ensure it is not expired and will not
expire during use.
2.
Allow the refrigerated ABC to equilibrate to room temperature prior to rst use.
Do not remov
e the seal until you have completed step 5.
3.
Verify that the seal is intact. Do not use if buer level is too low or seal has been
compromised. A ll tolerance of ±1 mm is acceptable.
Install the anode
buffer container
(ABC)
Chapter 9 Maint
ain the Instrument
Clean the instrument
9
3500/3500xL Genetic Analyzer User Guide—Data Collection Software v3.1
229