b.
Specify the order of the peaks in the calibration standard you are using. Use
the internal order number of the dye based on the dyes selected.
IMPORTANT! The Calibration Peak Order elds do not correspond to the
dye colors displayed above the Calibration Peak Order elds.
Example 1—If the order of the peaks in the calibration standard you are
using is Orange, Red, Yellow, Blue, Green, Purple, specify for
Calibration Peak Order: 6 (Orange), 4 (Red), 3 (Yellow), 1 (Blue), 2
(Green), 5 (Purple).
Example 2—If the order of the peaks in the calibration standard you are
using is Orange, Red, Yellow, Blue, Green, specify for Calibration Peak
Order: 5 (Orange), 4 (Red), 3 (Yellow), 1 (Blue), 2 (Green).
Example 3—If the order of the peaks in the calibration standard you are
using is Orange, Red, Green, specify for Calibration Peak Order: 3
(Orange), 2 (Red), 1 (Green).
6.
Expand the Parameters section, then specify remaining seings.
7.
Click Save.
Size standards library
A size standard denes the sizes of known fragments. It is used to generate a
standard curve. The standard curve is used to determine the sizing of unknown
samples.
When you create a sizecalling (fragment) or QC (HID) protocol, you add a size
standard to the protocol. If you add this item from the library, a copy of the item is
added to the protocol, and can be modied independently from the original items
stored in the library. For information on how changes are tracked if auditing is
enabled, see “Audit action“ on page 205.
Size standard
o
verview
Chapter 7 Manage libr
ary resources
Size standards library
7
170
3500/3500xL Genetic Analyzer User Guide—Data Collection Software v3.1

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Applied Biosystems 3500 and is the answer not in the manual?

Applied Biosystems 3500 Specifications

General IconGeneral
Number of capillaries8
Capillary Length50 cm
LaserSolid-state laser
Instrument TypeCapillary electrophoresis system
TechnologyCapillary Electrophoresis (CE)
Weight95 kg
Power Requirements100–240 V

Related product manuals