A spectral calibration can share capillary information:
Between injections – If a capillary in an injection does not meet the spectral
Quality V
alue and Condition Number limits shown on page 105, the software
automatically uses the information from that capillary in a dierent injection.
Within an injection – If a capillary in an injection does not meet the spectral
Quality Value and Condition Number limits shown on page 105 and the Allow
Borrowing option is selected, the software can also use the information from a
capillary to the left or the right of that capillary, if the values are higher than those
for that capillary in a dierent injection.
Spectral calibration with Borrowing disabled
When Borrowing is disabled, all capillaries must pass (meet the spectral Quality Value
and Condition Number limits) for the calibration to pass.
Injection 1
The software evaluates the Quality Value and Condition Number of
all c
apillaries.
If all capillaries pass, the calibration is complete, and injections 2
and 3 are not performed.
If any capillaries fail, injection 2 is performed.
Injection 2
The software evaluates the Quality Value for each capillary across
injections 1 and 2 and uses the information from the capillary with
the highest Quality Value.
If all capillaries now pass, the calibration is complete and injection
3 is not performed.
If the same capillary fails in both injection 1 and 2, injection 3 is
performed.
Injection 3
The software evaluates the Quality Value for each capillary across
injections 1, 2, and 3 and the information from the capillary with
the highest Quality Value.
If all capillaries now pass, the calibration passes.
If the same capillary fails in injection 1, 2, or 3, the calibration
fails.
Spectral calibration with Borrowing enabled
When Borrowing is enabled, all capillaries hav
e to pass (meet the spectral Quality
Value and Condition Number limits) within the borrowing limits:
8-capillary instruments – One adjacent-capillary borrowing event allowed
24-capillary instruments – Up to three adjacent-capillary borrowing events
allowed (the number of allowed borrowing events can be decreased in
Preferences).
Capillary
information
sharing
Chapter 6 Run c
alibrations and install checks
Run a spectral calibration
6
116
3500/3500xL Genetic Analyzer User Guide—Data Collection Software v3.1

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Applied Biosystems 3500 and is the answer not in the manual?

Applied Biosystems 3500 Specifications

General IconGeneral
Number of capillaries8
Capillary Length50 cm
LaserSolid-state laser
Instrument TypeCapillary electrophoresis system
TechnologyCapillary Electrophoresis (CE)
Weight95 kg
Power Requirements100–240 V

Related product manuals