The software identies a borrow
ed capillary with an arrow pointing from the
capillary from which the data is borrowed.
Injection 1
The software evaluates the Quality Value and Condition Number of
all c
apillaries.
If all capillaries pass, the calibration is complete, and injections 2
and 3 are not performed.
If any capillaries fail, the software borrows from an adjacent
capillary.
If, after borrowing, >1 or > 3 capillaries fail, injection 2 is
performed.
Injection 2
The software evaluates the quality values between adjacent
capillaries in injection 2 and for each capillary across injections 1
and 2 and uses the information with the highest Quality Value for
each capillary.
If all capillaries pass, the calibration is complete and injection 3 is
not performed.
If, after borrowing, >1 or > 3 capillaries from injection 1 or 2 do not
pass, injection 3 is performed.
Injection 3
The software evaluates the quality values between adjacent
capillaries in injection 3 and for each capillary across injections 1,
2, and 3, then uses the information with the highest Quality Value
for each capillary.
If all capillaries now pass, the calibration passes.
If after borrowing, >1 or > 3 capillaries from injection 1, 2, or 3 do
not pass, the calibration fails.
Chapter 6 Run calibrations and install checks
Run a spectral calibration
6
3500/3500xL Genetic Analyzer User Guide—Data Collection Software v3.1
117