Applied Biosystems 3500 - Replenish Consumables; Ensure Proper Installation of CBC Septa

Applied Biosystems 3500
326 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
If any consumables are expired or if buer ll lev
el is too low, replenish the
consumables as described in:
“Replenish polymer or change polymer type“ on page 232
IMPORTANT! Wear gloves while handling polymer, the capillary array, septa,
ABC, or CBC.
“Install the anode buer container (ABC)“ on page 229
“Install the cathode buer container (CBC)“ on page 230
“Fill capillary array with fresh polymer“ on page 234
“Install or change the capillary array“ on page 235
When you install the CBC buer septa, press rmly to seat the septa.
IMPORTANT! Look at the CBC from the side and ensure that there is no gap betw
een
the container and the lip of the septum.
Replenish
consumables
Ensure proper
installation of CBC
septa
Chapter 2 Start the system
Check system status in the Dashboard
2
3500/3500xL Genetic Analyzer User Guide—Data Collection Software v3.1
35