Restart the instrument and the computer
When to use this procedure:
If communication errors are displayed
If the front panel indicator is blinking red
At the end of spatial calibration, if Accept/Reject buons are dimmed
If maintenance wizards are taking longer than expected
If software operations are taking longer than expected
When you are instructed to restart the instrument and the computer:
1.
Exit the 3500 Series Data Collection Software 3.1.
2.
Power o the computer.
3.
Make sure the instrument door is closed, then power o the instrument.
4.
When the computer is completely powered o, wait 60 seconds, then power on
the computer. Wait until the Windows
login screen is displayed. Do not log in.
5.
Power on the instrument and wait until the green status light on the front panel
is on and not ashing before proceeding.
6.
Log in to Windows
operating system.
7.
Look in the Windows
taskbar at the boom right of the desktop and make sure
the Server Monitor icon is displayed. If it is not, go to “Software troubleshooting
— general“ on page 262.
8.
Start the 3500 Series Data Collection Softw
are 3.1.
Appendix A T
roubleshoot
Restart the instrument and the computer
A
3500/3500xL Genetic Analyzer User Guide—Data Collection Software v3.1
251

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Applied Biosystems 3500 and is the answer not in the manual?

Applied Biosystems 3500 Specifications

General IconGeneral
Number of capillaries8
Capillary Length50 cm
LaserSolid-state laser
Instrument TypeCapillary electrophoresis system
TechnologyCapillary Electrophoresis (CE)
Weight95 kg
Power Requirements100–240 V

Related product manuals