Review results
For sequencing data, review sample quality as described in “Review traces“ on
page 86.
For fragment analysis or HID data, review sample quality as described in
“Review sample quality“ on page 94.
If you change the order of columns before exporting a sizing table, Dye/Sample
Peak column values are split in to two separate columns, the column name
containing the Sample Peak values is incorrectly named, and the column headers
after Dye/Sample are shifted one column to the right. You can edit the exported
le: Cut/paste column headers one column to the right, enter correct headers for
the Dye and Sample Peak columns.
Appendix E Limitations
Review results
E
294
3500/3500xL Genetic Analyzer User Guide—Data Collection Software v3.1