Dye Set J6 created from Matrix Standard Set DS-36
To export spectral calibration results:
1.
Click Export Spectral Calibration Results.
2.
Specify an export le name and location, then click Save.
The export le contains the following results:
Capillary Number
Condition Number
Scan Number
Borrowed From Capillary
Quality V
alue
Peak Height
Reason For Failure
Run From Injection
Note: Spatial and spectral calibration reports include the date on which a capillary
array is installed for the rst time on the instrument. Install standard reports use the
most recent install date if a capillary array was removed and re-installed on the
instrument.
1.
Click
V
iew Report.
2.
(Optional) In the Report screen, click toolbar options to manipulate the report.
Place the mouse pointer over an item for a description of the item.
3.
To print the report, click
Print.
Export spectral
c
alibration results
View and print a
c
alibration report
Chapter 6 Run c
alibrations and install checks
Run a spectral calibration
6
3500/3500xL Genetic Analyzer User Guide—Data Collection Software v3.1
119

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Applied Biosystems 3500 and is the answer not in the manual?

Applied Biosystems 3500 Specifications

General IconGeneral
Number of capillaries8
Capillary Length50 cm
LaserSolid-state laser
Instrument TypeCapillary electrophoresis system
TechnologyCapillary Electrophoresis (CE)
Weight95 kg
Power Requirements100–240 V

Related product manuals