2.
In the Re-injection dialog box, select options, then click OK.
Note: Sample data les for each re-injection can be saved in a separate folder in
the results group folder if specied in the results group.
If you select a protocol other than the original, the software:
Creates a copy of the assay specied for the re-injected well (Original_Assay-1).
Adds the new or modied instrument protocol to Original_Assay-1.
Assigns Original_Assay-1 to the re-injected well only.
Saves the plate (the software does not save the copy of the assay to the library).
If the Injection Number aribute is selected for display in the plate view, the number
of the original injection and the re-injection are shown.
1
2
1
Sample 1 selected for re-injection
2
Re-injection number listed for all samples in the re-injection
If you select a
pr
otocol other
than the original
How re-injections
are displayed in
the plate view
Chapter 3 Set up and run
Check sequence or sample quality and re-inject samples
3
70
3500/3500xL Genetic Analyzer User Guide—Data Collection Software v3.1

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Applied Biosystems 3500 and is the answer not in the manual?

Applied Biosystems 3500 Specifications

General IconGeneral
Number of capillaries8
Capillary Length50 cm
LaserSolid-state laser
Instrument TypeCapillary electrophoresis system
TechnologyCapillary Electrophoresis (CE)
Weight95 kg
Power Requirements100–240 V

Related product manuals