Instrument
On-instrument supported limits
Lower of:
Guidelines
8-capillary 14 days, 240 injections, or expiry date The buffer has been verified for use for
up to 14 days on the instrument.
The software displays a warning
message when a usage limit is met and
allows you to continue running. Before
doing so, see “Important notice
regarding use of consumables that
exceed supported limits“ on page 22.
24-capillary 14 days, 100 injections, or expiry date
Polymer
Cat. No.
Description Storage
A26070 POP-4
™
Polymer (96‑sample) 2–8°C
4393715 POP-4
™
Polymer (384‑sample)
[1]
4393710 POP-4
™
Polymer (960‑sample)
[1]
A26071 POP-6
™
Polymer (96‑sample )
4393717
POP-6
™
Polymer (384‑sample)
4393712 POP-6
™
Polymer (960‑sample)
A26073 POP-7
™
Polymer (96‑sample)
4393708 POP-7
™
Polymer (384‑sample)
4393714 POP-7
™
Polymer (960‑sample)
[1]
The polymer has been validated for HID applications.
IMPORTANT! For the POP-4
™
and POP-7
™
P
olymers (Cat. Nos. A26070, 4393715,
4393710, A26073, 4393708, and 4393714), the on-instrument supported limit is 14 days
only when the instrument operating temperature is 15–25°C. When the instrument
operating temperature is > 25°C, the supported limit is 7 days.
For the POP-6
™
Polymers (Cat. Nos. A26071, 4393717, and 4393712), the
on-instrument supported limit is 14 days when the instrument operating temperature
is 15–30°C.
Chapter 1 Instrument and software description
Instrument consumables handling, usage limits, and expiration
1
20
3500/3500xL Genetic Analyzer User Guide—Data Collection Software v3.1