5.
If the data for the required number of capillaries do not meet the criteria above
(at least 7 capillaries for 8-capillary instruments, at least 22 capillaries for
24-capillary instruments):
a.
(Optional) If you want to generate a report for the failed calibration, click
View Report before y
ou click Reject Results. To save the report
electronically, select CutePDF as the printer.
b.
Click Reject Results. For troubleshooting information, see “Sequencing
install standard troubleshooting“ on page 274.
IMPORTANT! If y
ou reject results, the spectral calibration is not saved.
Select History V
iew, then select an install standard to view the associated install
check information.
Example
sequencing ins
tall
check results
View previously
run ins
tall
standards
Chapter 6 Run c
alibrations and install checks
Run a Sequencing install check
6
3500/3500xL Genetic Analyzer User Guide—Data Collection Software v3.1
129