EasyManua.ls Logo

Applied Biosystems 3500 - QC Protocols Library (Primary Analysis-HID); QC Protocol Overview; Create a New QC Protocol

Applied Biosystems 3500
326 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
QC protocols library (primary analysis—HID)
A QC protocol is the required primary analysis protocol for HID applications. A QC
protocol denes peak detection, sizing, and quality values.
When you create an HID assay, you add a QC protocol to the assay. If you add this
item from the library, a copy of the item is added to the assay, and can be modied
independently from the original item stored in the library. For information on how
changes are tracked if auditing is enabled, see “Audit action“ on page 205.
If factory-provided QC protocols do not suit your needs, you can create new QC
protocols:
1.
Access the QC Protocols library.
2.
Click Create.
3.
In the Analysis Seings tab of the Create New QC Protocol dialog box
(Figure 27), specify seings (“QC protocol—Analysis seings“ on page
187).
4.
Click QC Seings. In the QC Seings tab of the Create New QC Protocol dialog
box (Figure 28), specify seings (“QC protocol—QC seings“ on page 190).
QC protocol
o
verview
Create a new QC
protocol
Chapter 7 Manage library resources
QC protocols library (primary analysis—HID)
7
3500/3500xL Genetic Analyzer User Guide—Data Collection Software v3.1
185

Table of Contents

Related product manuals