QC protocols library (primary analysis—HID)
A QC protocol is the required primary analysis protocol for HID applications. A QC
protocol denes peak detection, sizing, and quality values.
When you create an HID assay, you add a QC protocol to the assay. If you add this
item from the library, a copy of the item is added to the assay, and can be modied
independently from the original item stored in the library. For information on how
changes are tracked if auditing is enabled, see “Audit action“ on page 205.
If factory-provided QC protocols do not suit your needs, you can create new QC
protocols:
1.
Access the QC Protocols library.
2.
Click Create.
3.
In the Analysis Seings tab of the Create New QC Protocol dialog box
(Figure 27), specify seings (“QC protocol—Analysis seings“ on page
187).
4.
Click QC Seings. In the QC Seings tab of the Create New QC Protocol dialog
box (Figure 28), specify seings (“QC protocol—QC seings“ on page 190).
QC protocol
o
verview
Create a new QC
protocol
Chapter 7 Manage library resources
QC protocols library (primary analysis—HID)
7
3500/3500xL Genetic Analyzer User Guide—Data Collection Software v3.1
185

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Applied Biosystems 3500 and is the answer not in the manual?

Applied Biosystems 3500 Specifications

General IconGeneral
Number of capillaries8
Capillary Length50 cm
LaserSolid-state laser
Instrument TypeCapillary electrophoresis system
TechnologyCapillary Electrophoresis (CE)
Weight95 kg
Power Requirements100–240 V