5.
Click Save.
IMPORTANT! The default values in the QC protocol templates (other than peak
amplitude threshold values) have been optimized for each kit. You must
optimize and validate peak amplitude threshold values during internal HID
validation. If you modify other seings, ensure that the size standard is
accurately detected and sized with the new seings.
IMPORTANT! Normalization is not applied to samples with Size Quality
ags. The 3500 Series Data Collection Softw
are 3.1 does not support re-analyzing
data with new seings.
Figure 27 Create New QC Protocol—Analysis Settings
Chapter 7 Manage library resources
QC protocols library (primary analysis—HID)
7
186
3500/3500xL Genetic Analyzer User Guide—Data Collection Software v3.1