Review sizing
The Sizing Table View displays:
For fragment samples—All dyes
For HID samples—Size standard dye only (orange or red)
1.
Select samples in the samples table to display the plots.
2.
In the sizing table, click the Table Seings buon, then specify the columns to
show or hide.
3.
Filter the table as needed.
4.
Double-click column headers to sort columns. Multicolumn sorting is supported
(see “Sort by one or multiple columns“ on page 76).
5.
Selecting rows in the sizing table, then click Label Selected Peaks.
1.
In the plot view toolbar
, deselect all dye colors except the size standard dye color
(red or orange).
2.
In the sizing table, select the size standard peaks of interest.
3.
Click Label Selected Peaks to label the size standard peaks in the plot view.
Note: If labels are not displayed, click
(Plot Seings) in the plot view toolbar
,
then select Show Labels in the Labels tab. Click Save to Preferences to retain this
seing.
4.
Ensure that all size standard peaks are present and correctly labeled.
Sort by one or
multiple columns
Set up the sizing
t
able
Examine the size
s
tandard plot
Chapter 5 Re
view fragment/HID analysis results
Review sizing
5
100
3500/3500xL Genetic Analyzer User Guide—Data Collection Software v3.1

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Applied Biosystems 3500 and is the answer not in the manual?

Applied Biosystems 3500 Specifications

General IconGeneral
Number of capillaries8
Capillary Length50 cm
LaserSolid-state laser
Instrument TypeCapillary electrophoresis system
TechnologyCapillary Electrophoresis (CE)
Weight95 kg
Power Requirements100–240 V