Specifications
Table 16 RFID read/write unit specifications
Component Specification
RFID read/write unit
Ultra-Compact Proximal-Type RFID Reader / Writer
Model ASI4000-98-BS1
Manufactured by ART Technology Co., Ltd.
RF frequency 13.56 MHz
RF output power 60 mW
RFID tags Texas Instruments RI-I03-112A-03 tags, tested by the manufacturer
to reliably read and write 100,000 times with zero data loss and retain
written data for more than 10 years
Effective range between RFID tag and
internal RFID read/write units
ABC tag: 3 cm
CBC tag: 4 cm
Capillary tag: 3 cm
Polymer tag: 3 cm
Typical use range between RFID tag and
internal RFID read/write units
0.5 cm
Minimum separation distance of the
instrument from external RFID read/write
units
10 cm
Minimum separation distance of the
instrument from other wireless technologies
3 feet
Wireless security
RFID tag read/write/re-read with checksum
Password access for use of software
Base-64 encoding of data between the instrument and the
computer