4.
Click a row in the ag table, then click the Sample tab in Instrument Run Views
to display the associated data in the Sample view.
A re-injection physically re-injects all samples in the capillary array. You can specify
whether to collect data for all or only selected samples in the array
.
By default, you can specify a re-injection before the run completes. To allow re-
injections after a run is complete, set the Pause After Last Injection preference (see
“User preferences“ on page 37).
1.
Select the injections or wells to re-inject:
Note:
Re-inject is dimmed if you select an injection that contains more than
one results group, or if y
ou select ags in the ags table that correspond to
samples with dierent results groups. To enable
Re-inject, select samples that
specify the same results group.
To collect data for all wells in an injection
1. Sel
ect the injection in the injection
list.
2. Click
Re-inject.
To collect data for only specific w
ells
Note: You can also re-inject specific
samples in Review Results.
1. Select the injection.
2. Select in the array view the capillary
that corresponds to the well or
sample of interest (see “Array
view“ on page 81).
3. Click
Re-inject.
To collect data for only samples that
c
ontain flags
1. Select the samples in the flag table
(see “Check sequence or sample
quality“ on page 67).
2. Click
Re-inject.
Note: If y
ou are running an HID plate, see “Re-inject HID allelic ladder
samples“ on page 71.
Reinject sampl
es
from the Monitor
Run screen
Chapter 3 Set up and run
Check sequence or sample quality and re-inject samples
3
3500/3500xL Genetic Analyzer User Guide—Data Collection Software v3.1
69

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Applied Biosystems 3500 and is the answer not in the manual?

Applied Biosystems 3500 Specifications

General IconGeneral
Number of capillaries8
Capillary Length50 cm
LaserSolid-state laser
Instrument TypeCapillary electrophoresis system
TechnologyCapillary Electrophoresis (CE)
Weight95 kg
Power Requirements100–240 V

Related product manuals