5-38 Starling Monitor - User Guide
Performing a Dynamic Assessment Test
The Stage 1 wizard opens.
5. As instructed, place the patient in the semi-recumbent position, then tap Next.
Keep the patient still in this position for 3 mins.
During this time, the system will obtain a Baseline for measurement.
While the Baseline measurement is in progress, the Dynamic display shows a graph
being drawn slowly from left to right, as well as a 3-minute countdown (min/secs). See
the example in Figure 5-34.