©2020 Baxter International Inc. 1- 15
Guidelines for Safe Handling of the Starling Monitor
Environmental Requirements
The manufacturer’s environmental recommendations for operation and storage of the
Starling Monitor are shown in Table 1-4.
The manufacturer’s recommendations for storage temperature of the Starling Sensors are
shown in Table 1-5.
Table 1-4 Recommended Environmental Conditions - Starling Monitor
Condition
Ambient
Temperature
Relative Humidity
(non-condensing)
Atmospheric
Pressure (hPa)
Operation, Storage,
and Transportation
5°C — 50°C
15% — 85% 700 — 1060
41°F — 122°F
Table 1-5 Recommended Temperature Conditions - Starling Sensors
Condition
Ambient
Temperature
Storage
5°C—30°C
41°F—86°F