EasyManua.ls Logo

baxter Starling - Appendix B - System Specifications and Technical Data; RF Wireless Communications Immunity Specifications

baxter Starling
266 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
©2020 Baxter International Inc. B- 13
Technical Information
Immunity to Proximity Fields from RF Wireless Communications Equipment
Table B-11 Test Specifications for Enclosure Port Immunity to RF Wireless Communications
Test
Frequency
(MHz)
Band
(MHz)
Service Modulation
Maximum
Power
(W)
Distance
(m)
IMMUNITY
TEST LEVEL
(V/m)
Compliance
Level
385 380-390 TETRA 400
Pulse modulation
18 Hz
1.8 0.3 27 27
450 430-470
GMRS 460
FRS 460
FM
± 5 kHz deviation
1 kHz sine
5 0.3 28 28
710
704-787 LTE Band 13, 17
Pulse modulation
217 Hz
0.2 0.3 9 9745
780
810
800-960 GSM 800/900
Pulse modulation
18 Hz
2 0.3 28 28870
930
1720
1700-1990
GSM 1800;
CDMA 1900;
GSM 1900;
DECT;
LTE Band
1, 3, 4, 25;
UNITS
Pulse modulation
217 Hz
2 0.3 28 28
1845
1970
2450 2400-2570
Bluetooth,
WLAN,
802.11 b/g/n,
RIFD 2450,
LTE Band
Pulse modulation
217 Hz
2 0.3 28 28
5240
5100-5800
WLAN,
802.11 a/n
Pulse modulation
217 Hz
0.2 0.3 9 95500
7585

Table of Contents