©2020 Baxter International Inc. C- 17
SpO
2
Module (Optional)
SpO
2
Error Code List and Definitions
Table C-6 SpO
2
Module - Error Code List and Definitions
Error Code Recommended Corrective Action
SpO2 Low signal, please check sensor
Adjustment.
Readjust the SpO
2
sensor.
If problem persists, replace the SpO
2
sensor.
If problem still persists, contact Baxter Service Center:
email:
starling_rma@baxter.com
Tel (worldwide): +1 360-828-8685
SpO2 System is being Reset.
Wait until the SpO
2
measurements are initiated.
If after 3 minutes the SpO
2
measurements are not updated, service may be
required. Contact Baxter Service Center:
email:
starling_rma@baxter.com
Tel (worldwide): +1 360-828-8685
SpO2 Module Error, please contact
Cheetah Medical.
Service is required. Contact Baxter Service Center:
email: starling_rma@baxter.com
Tel (worldwide): +1 360-828-8685
SpO2 sensor malfunction, please replace
SpO2 Sensor.
Stop the SpO
2
Module.
Replace the SpO
2
sensor.
Start the SpO
2
Module.