B-12 Starling Monitor - User Guide
Technical Information
Recommended Separation Distances
Table B-10 Recommended Separation Distances
Recommended Separation Distances between portable and mobile RF
communications equipment and the Starling Monitor
The Starling Monitor is intended for use in an electromagnetic environment
in which radiated RF disturbances are controlled.
The user and/or installer of the unit can help prevent electromagnetic interference by maintaining a minimum
distance between portable and mobile RF communications equipment (transmitters) and the Starling Monitor,
according to the maximum output power of the equipment, as recommended below.
Rated maximum output
power of the transmitter
Watts (W)
Separation Distance according to frequency of transmitter
in Meters (m)
150 kHz to 80 MHz
d = 1.17 P
80 MHz to 800 MHz d
= 1.17 P
800 MHz to 2.5 GHz
d = 2.3 P
0.01 0.12 0.12 0.23
0.1 0.37 0.37 0.73
1 1.17 1.17 2.3
10 3.7 3.7 7.3
100 11.7 11.7 23