Table B-9 Guidance and Manufacturer’s Declaration
Electromagnetic Immunity for the Starling Monitor that is Not Life-Supporting
Immunity
Test
Basic EMC
Standard
or Test Method
Immunity Test Levels
Professional Health Care
Facility Environment
Compliance Level
Conducted RF IEC 61000-4-6
3 V
0.15 MHz to 80 MHz
6 V
in ISM bands
between 0.15 MHz and 80 MHz
80% AM at 1 kHz
3 V
0.15 MHz to 80 MHz
6 V
in ISM bands
between 0.15 MHz and 80 MHz
80% AM at 1 kHz
Radiated RF IEC 61000-4-3
3 V/m
80 MHz to 2.7 GHz
80% AM at 1 kHz
3 V/m
80 MHz to 2.7 GHz
80% AM at 1 kHz