C-10 Starling Monitor - User Guide
NIBP Module
Error Code List and Definitions
When an NIBP error occurs, measurement values will still be reported.
These measurement values should be disregarded when an error code is reported.
Table C-3 NIBP Module - Error Code List and Definitions
Error Code Recommended Corrective Action
Artifact / Erratic Oscillometric Signal.
The patient may have been moving too much.
Check that the cuff is in the correct positions.
Check that the correct size cuff is being applied.
Exceeded retry count.
The patient may have been moving too much.
Check that the cuff is properly tightened.
Check that the cuff is in the correct position.
Check that the correct size cuff is being applied.
Check that there is no excessive clothing between the arm
and the cuff.
Exceeded measurement time limit.
The patient may have been moving too much.
Check that the cuff is properly tightened.
Check that the cuff is in the correct position.
Check that the correct size cuff is being applied.
Check that there is no excessive clothing between the arm
and the cuff.
Pneumatic Blockage.
Check that the hose has no sharp bends or is pinched.
Check that the patient is not lying on the cuff.