©2020 Baxter International Inc. C- 9
NIBP Module
Pressure Transducer Accuracy
±3mmHg between 0 mmHg and 300 mmHg for operating conditions between
0°C and 50°C
Recommended Frequency of Pressure Transducer
Calibration
The pressure transducer calibration should be verified annually.
Operating Conditions 0°C to 50°C (32°F to 122°F), 15% to 95% non-condensing humidity
Storage Conditions -20°C to 65°C (-4°F to 149°F), 15% to 90% non-condensing humidity
Altitude Measurement accuracy is not affected by altitude
Startup Initialization Period 7 seconds
Minimum time between commands 50 milliseconds
Patient Safety Internal operating software ensures that:
Maximum initial inflation time is limited to 75 seconds
Duration of blood pressure reading is limited to:
130 seconds (Adult/Pediatric mode)
75 seconds (Neonate mode)
Additional redundant safety circuitry oversees normal operation and will
override to abort a reading if:
Cuff pressure exceeds 300 mmHg (Adult & Pediatric modes) or 150mmHg
(Neonate mode) at any time the cuff has been inflated for 180 seconds
(Adult & Pediatric modes) or 90 seconds (Neonate mode)
The module meets all relevant parts of the following Safety Standards:
IEC60601-1:2005 3rd Edition
IEC/EN60601-2-30:1999/2000
IEC 80601-2-30:2009
ISO 81060-2:2009
Table C-2 NIBP Module Performance (continued)
Parameter Details