©2020 Baxter International Inc. B-1
Appendix B - System Specifications
and Technical Data
System Performance Specifications and Accuracy
Monitored Parameters
Table B-1 Normal Hemodynamic Parameters
Parameter Equation
Normal Adult Range
a,b,c,d
a. Cecconi M, Parsons AK, Rhodes A. What is a Fluid Challenge?
Curr Crit Care 2011; (17)3;290-5
.
b. Monnet X. Teboul JL. Passive Leg Raising.
Intensive Care Med 2008; 34;659-63
.
c. Marik P. Monnet X. Teboul JL. Hemodynamic Parameters to Guide Fluid Therapy.
Annals of Intensive Care. 2011
.
d. Sramek BB. Systematic Hemodynamics and Hemodynamic Management 2002;
Instapublisher.com ISBN 1-59196-04600
Stroke Volume (SV) CO/HR x 1000 60 – 100 mL/beat
Stroke Volume Index (SVI) SV/BSA
33 – 47 mL/beat/m
2
Stroke Volume (SVI)
Change in SV after
Dynamic Assessment
>
10% - Likely to be Fluid Responsive
< 10% - Unlikely to be Fluid Responsive
Cardiac Output (CO) HR x SV/1000 4.0 – 8.0 L/min
Cardiac Index (CI) CO/BSA
2.5 – 4.0 L/min/m
2
Total Peripheral Resistance (TPR) 80 x (MAP)/CO
800 – 1200 dynes • sec/cm
–5
Total Peripheral Resistance Index (TPRI) 80 x (MAP)/CI
1970 – 2390 dynes • sec/cm
–5
/m
2