B-10 Starling Monitor - User Guide
Technical Information
Table B-8 Guidance and Manufacturer’s Declaration - Electromagnetic Immunity
for the Starling Monitor
The Starling Monitor is designed to be used in the electromagnetic environment specified below.
The user and/or installer of the unit must ensure that it is used in such an environment
.
Immunity
Test
IEC 60601-1-2
Test Level
Compliance
Level
Recommended Electromagnetic Environment
Electrostatic
discharge (ESD)
IEC 61000-4-2
± 8 kV for contact
± 2 kV, ± 4 kV, ± 8 kV,
± 15 kV air
± 8 kV contact
± 2 kV, ± 4 kV, ± 8 kV,
± 15 kV air
Floors should be wood, concrete or ceramic tile. If floors are
covered with synthetic material, the relative humidity should be
at least 30%.
Electrical fast
transient/ burst
IEC 61000-4-4
± 2 kV for power supply
line
± 1 kV for input / output
lines
± 2 kV for power supply
line
± 1 kV for input / output
lines
Mains power quality should be that of a typical public low-
voltage power supply network that supplies buildings used for
domestic purposes, commercial or hospital, clinic
environment.
Surge
IEC 61000-4-5
± 0.5 kV, ± 1 kV
line to line
± 0.5 kV, ± 1 kV, ± 2 kV
line to earth
± 0.5 kV, ± 1 kV
line to line
± 0.5 kV, ± 1 kV, ± 2 kV
line to earth
Mains power quality should be that of a typical public low-
voltage power supply network that supplies buildings used for
domestic purposes, commercial or hospital, clinic
environment.
Voltage dips,
short
interruptions and
voltage
variations on
power supply
input lines
IEC 61000-4-11
0% U
T
for 0.5 cycle
0% U
T
for 1 cycle
and
70% U
T
for 25 cycles
Single phase: at 0
o
0% U
T
for 250 cycles
0% U
T
for 0.5 cycle
0% U
T
for 1 cycle
and
70% U
T
for 25 cycles
Single phase: at 0
o
0% U
T
for 250 cycles
Mains power quality should be that of a typical public low-
voltage power supply network that supplies buildings used for
domestic purposes, commercial or hospital, clinic
environment.
If the user of the Starling Monitor requires continued operation
during power mains interruptions, it is recommended that the
Starling be powered from a separate power supply (UPS,
etc
.).
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
3 A/m
50 Hz
3 A/m
50 Hz
Mains power quality should be that of a typical public low-
voltage power supply network that supplies buildings used for
domestic purposes, commercial or hospital, clinic
environment.
NOTE: U
T
is the AC mains voltage prior to application of the test level.