C-8 Starling Monitor - User Guide
NIBP Module
NIBP Module Performance
Table C-2 NIBP Module Performance
Parameter Details
Method of Measurement Oscillometric. Diastolic values correspond to Phase 5 Korotkoff sounds.
Blood Pressure Range
mmHg
Systolic
ADULT 40 - 260
PEDIATRIC 40 - 230
NEONATE 40 - 130
MAP
ADULT 26 - 220
PEDIATRIC 26 - 183
NEONATE 26 - 110
Diastolic
ADULT 20 - 200
PEDIATRIC 20 - 160
NEONATE 20 - 100
Pulse Rate Range
30 to 220 BPM (Beats Per Minute)
Pulse Rate Accuracy
± 2% or ± 3 BPM, whichever is greater
Cuff Deflate Rate
Deflation step size varies with heart rate, cuff pressure and cuff volume.
Initial Inflation Pressure
mmHg
ADULT
160 (default)
Variable from 120 to 280
PEDIATRIC
140 (default)
Variable from 80 to 280
NEONATE
90 (default)
Variable from 60 to 140
Clinical Accuracy
Meets accuracy requirements of the following Safety Standards:
IEC 80601-2-30:2009
ISO 81060-2:2013.