©2020 Baxter International Inc. C- 11
NIBP Module
Inflate Timeout, Air Leak or Loose Cuff.
Check that the hose is connected to the system and the cuff.
Check that the cuff is properly tightened.
Check that the cuff is in the correct position.
Check that the correct size cuff is being applied.
Check that the cuff is not leaking air.
Check that the hose connections are not damaged or loose.
Safety Timeout.
Check the patient.
Check that the cuff is in the correct position.
The patient may have been moving too much.
Take another BP reading.
Cuff Overpressure.
Check that the correct size cuff is being applied.
Check that the hose has no sharp bends or is pinched.
Check that the cuff is in the correct position.
Check that the patient is not lying on the cuff.
Power supply out of range or other
hardware problem.
Service may be required. Contact Baxter Service Center:
email: starling_rma@baxter.com
Tel (worldwide): +1 360-828-8685
Permission problem.
Service may be required. Contact Baxter Service Center:
email:
starling_rma@baxter.com
Tel (worldwide): +1 360-828-8685
Transducer out of range.
Service may be required. Contact Baxter Service Center:
email: starling_rma@baxter.com
Tel (worldwide): +1 360-828-8685
Table C-3 NIBP Module - Error Code List and Definitions (continued)
Error Code Recommended Corrective Action