C-12 Starling Monitor - User Guide
NIBP Module
ADC out of range.
Service may be required. Contact Baxter Service Center:
email:
starling_rma@baxter.com
Tel (worldwide): +1 360-828-8685
EEPROM calibration data failure.
If possible, perform calibration.
Service may be required. Contact Baxter Service Center:
email:
starling_rma@baxter.com
Tel (worldwide): +1 360-828-8685
Error Measuring NIBP – Critical NIBP
module error.
Service may be required. Contact Baxter Service Center:
email: starling_rma@baxter.com
Tel (worldwide): +1 360-828-8685
Device Busy – NIBP device is not
responding as it is in the middle of an
activity.
Retry to obtain an NIBP reading.
If this error persists, restart the unit and try again to obtain an
NIBP reading.
If you still encounter the error message after restarting the unit, service may be
required. Contact Baxter Service Center:
email:
starling_rma@baxter.com
Tel (worldwide): +1 360-828-8685
Table C-3 NIBP Module - Error Code List and Definitions (continued)
Error Code Recommended Corrective Action