5-40 Starling Monitor - User Guide
Performing a Dynamic Assessment Test
As soon as the countdown reaches zero, Baseline measurement is complete.
The system sounds a single beep; the Stage 2 wizard opens.
Note
In the event that the Baseline results are considered by the system to be unstable, a warning
message is displayed. The user has the option of nevertheless continuing to the Challenge stage, or
restarting the Test again from the Baseline stage.
6. As instructed, place the patient in supine position, then elevate the legs at an angle of
45 degrees.
Note
To support the patient’s legs comfortably, use of the Lift (optional accessory) is recommended.
See Figure 2-16 on page 2-28.
7. Tap Next.
The Challenge commences.
8. Keep the patient still in this position for 3 mins.
Figure 5-35 PLR Test - Stage 2 Wizard