C-14 Starling Monitor - User Guide
SpO
2
Module (Optional)
Cleaning and maintenance instruction for the SpO
2
sensor are described in the specific SpO
2
sensor user manual which is supplied with the sensor.
The SpO
2
sensor is connected to the Starling Monitor via the dedicated SpO
2
cable (illustrated
in Figure 2-12 on page 2-24) that connects to the dedicated connection on the side of the
Monitor (shown in Figure 2-8 on page 2-13).
Table C-4 NELLCOR SpO
2
Sensors - Accuracy Specifications
Type Sensor
Adults/Pediatric
Accuracy
(digits)
Neonates
Accuracy
a
(digits)
a.Neonatal Accuracy: When sensors are used on neonatal subjects as recommended, the specified
accuracy range is increased by ± 1 digit, as compared to adult usage, to account for the theoretical
effect on oximeter measurements of fetal hemoglobin in neonatal blood.
For example, OxiCliq N accuracy on neonates is ± 3.5 digits, rather than ± 2.5.
Reusable
D-YS (Infant to Adult) ± 3 ± 4
D-YS with D-YSE Ear Clip ± 3.5 N/A
D-YS with D-YSPD SpotClip ± 3.5 N/A
DS-100A ± 3 N/A
OXI-A/N ± 3 ± 4
OXI-P/I ± 3 N/A
Disposable
OxiCliq A ± 2.5 N/A
OxiCliq P ± 2.5 N/A
OxiCliq N (Adult) ± 2.5 ± 3.5
OxiCliq I ± 2.5 N/A