CAPINTEC, INC. CAPTUS
®
3000
Figure 10-28
Default Values frame
Administered to Standard Activity Ratio – This is the standard dilution factor,
which is 1000 for this procedure.
Sample Volume – This is the volume used for the samples counted in the well
counter, and should not exceed the useful counting dimensions of the well. A sample
volume of 4 ml is a good sample for the well detector provided with the CAPTUS
®
3000 system.
Note: The volume must be the same for both the patient’s sample and the standard
solution sample.
Counting Time – This is the duration for counting, in seconds, each time a
measurement is performed. The default value is set to 60 seconds.
Repetitions frame – This is the number of repetitions each time a measurement is made.
For two repetitions, the average value is used in calculations. The second repetition may
be skipped during the measurement procedure, if so desired.
10-20 LAB TESTS November 13