DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL
3 - 22 Section 3-5 - Preparing for Setup
3-6-5-1 Connect a Probe (cont’d)
5.) Position the Probe Cables in a way which the are not sweeping the floor.
3-6-5-2 Disconnect Probe
1.) Rotate the lock handle counter-clockwise to the horizontal position to unlock the connector.
(See Figure 3-11.)
2.) Remove the connector from the port.
3.) Ensure that the probe head is clean before placing the probe in its storage case, see section 10-5-
4 on page 10-9 for cleaning instructions.
Figure 3-11 Connect Probe
CAUTION
-
Take the following precautions with the probe cables:
- Keep the probe cables away from the wheels.
- Do not bend the probe cables.
- Do not cross cables between probes.
Probe Connectors
Parking Slot
for probe
connector
Locked position
Unlocked position