GE HEALTHCARE
DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
3 - 78 Section 3-9 - Connectivity Setup - Software v7.x (BT’08), v6.x, v5.x and v4.x
3-9-17 Query/Retrieve (Q/R) Setup for Software v7.x/v6.x/v5.x/v4.x
3-9-17-1 Overview
The Query/Retrieve function makes it possible to search for and retrieve DICOM data from a DICOM 
server for further analysis on the Vivid 7. 
NOTE: You may have to set up Vivid 7 as a destination on the server.
3-9-17-2 Query/Retrieve Setup on the Vivid 7 with Software 
v7.x/v6.x/v5.x/v4.x
1.) Press CONFIG (F2) and log on as adm, see 4-2-4 "Log On to the System as ‘ADM’" on page 4-12.
2.) Select Connectivity
3.) Select Dataflow
4.) Select Query Retrieve from the Name pull down menu
Figure 3-77   Select Query Retrieve
NAME PULLDOWN 
MENU
QUERY RETRIEVE