3 - 82 Section 3-9 - Connectivity Setup - Software v7.x (BT’08), v6.x, v5.x and v4.x
3-9-17-3 Query/Retrieve Verification
Follow the Steps Below to do a First Test (TCP-IP Ping) of the Connection
1.) Select Query retrieve and then select Properties.
2.) Select the “Smiley” button to Ping the server.
NOTE: The smiley checks if the remote server is accessible (ping). It is not a DICOM Echo (DICOM
ping), so it does not check AE title or port number.
3.) If the network connection to the server is OK, it will be illustrated by an smiling “Smiley” .
A sad Smiley indicates that the network connection is failing.
Typical causes:
- Network cable not connected.
- Configuration error(s).
4.) When ready, select OK to close the DICOM Query/Retrieve properties and save changes.
Figure 3-82 DICOM Query/Retrieve properties
SMILEY
(NETWORK
PING)
OK