GE HEALTHCARE
DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL
Chapter 3 - System Setup 3 - 81
3-9-17-2 Query/Retrieve Setup on the Vivid 7 with Software v7.x/v6.x/v5.x/v4.x (cont’d)
Change Search Criterias
It is possible to set up special Search Criterias for DICOM Query/Retrieve. In most cases you may leave
the Search Criterias as is, and skip this adjustment.
Follow the steps below to change the Search Criterias parameters:
1.) Select Search Criterias
2.) Select the correct tag from the Select Tag pull-down menu.
3.) If needed, type in the value.
NOTE: In this example, tag 00080060 Modality has been added to the list with value “US”. This may be
useful if the studies contain non-ultrasound images.
4.) Select Add to List
5.) Select OK to close the Search Criterias window.
Figure 3-81 Select Search Criterias
SEARCH CRITERIAS
ADD TO LIST
REMOVE
SELECT TAG