EasyManua.ls Logo

GE Vivid 7 - Page 229

GE Vivid 7
858 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
GE HEALTHCARE
DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL
Chapter 3 - System Setup 3 - 105
3-10-10-2 Query/Retrieve Setup on the Vivid 7 (cont’d)
Change Search Criterias
If is possible to set up special Search Criterias for DICOM Query/Retrieve. In most cases you may leave
the Search Criterias as is, and skip this adjustment.
Follow the steps below to change the Search Criterias parameters:
1.) Select Search Criterias
2.) Select the correct tag from the Select Tag pull-down menu.
3.) If needed, type in the value.
NOTE: In this example, tag 00080060 Modality has been added to the list with value “US”. This may be
useful if the studies contain non-ultrasound images.
4.) Select Add to List
5.) Select OK to close the Search Criterias window.
Figure 3-111 Select Search Criterias
SEARCH CRITERIAS
SELECT TAG
ADD TO LIST
REMOVE

Table of Contents

Related product manuals