GE HEALTHCARE
DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL
Chapter 3 - System Setup 3 - 123
3-11-9-3 Configuration of Modality Worklist Services (cont’d)
Configure the Search Criterias
NOTE: This configuration is required.
1.) Select the Search Criterias button.
2.) Select the criteria Modality and select the Remove button.
3.) Press the OK button.
Configure Search Criterias: Constrain Query by AE Title
NOTE: This configuration is optional. It should only be performed if the site is directing orders to the
Vivid 7 that is scheduled to perform the study.
Please contact the Mitra support personnel to verify that this step is needed.
1.) In the Search Criterias dialog box, select 00400001 Scheduled Station AE Title from the Select Tag
pull down menu.
2.) Enter the Vivid 7 computer name for the Value.
3.) Press the Add to List button.
4.) Press the OK button.
Configure Search Criterias: Constrain Query by MRN
NOTE: This configuration should be performed if the site will need to constrain the Worklist queries by
MRN (patient id)
1.) In the Search Criterias dialog box, select 00101000 Other Patient IDs from the Select Tag pull down
menu.
2.) Enter a 9 digit number for the Value.
3.) Select the Add to List button.
4.) Select the OK button.
Assign Dicom Worklist name to Dataflow
1.) Select the Dataflow tab.
2.) From the Dataflow name pull down menu, select the Dataflow which will be used.
3.) Remove the Dicom Worklist service currently used in the Dataflow.
4.) From the Select Service pull down menu, select the Gateway Modality Worklist service.
5.) Select the Add button.
6.) Ensure that the service is displaying the value In for the Dir parameter and that its role is Primary.