GE HEALTHCARE
DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL
4 - 60 Section 4-3 - Functional Checks
4-3-11-4 Check Invert, Tissue Tracking, Simultaneous and Color Maps
Table 4-24 TVI Mode Controls Checks
Step Task Expected Result(s)
5. Press INVERT
Enables the color scheme assigned to positive and
negative tissue velocities to be inverted.
Invert is available in live and cine replay.
6. Press TISSUE TRACKING
Starts Tissue Tracking mode.
7. Choose COLOR MAPS
Displays a drop down menu of different colorization
maps.
Use the trackball to select the desired map and press
Select to activate the map.