4-3-11-5 Check Tissue Priority, Frame Rate, Compress, Reject, Frequency, Lateral & Radial Average, 
Baseline and Power
 
Table 4-25    TVI Mode: Soft Menu Controls Check 
Step Task  Expected Result(s)
8. Adjust TISSUE PRIORITY 
Emphasizes either the color of the TVI or the greyscale 
tissue detail of the 2D image.
Tissue priority is available in both Live and Freeze.
9. Adjust FRAME RATE
 
Controls the line density. When adjusting frame rate, 
there is a trade off between spatial and temporal 
resolution.
10. Adjust COMPRESS
 
Controls the amount of color compression. The color 
bar is adjusted accordingly.
11. Adjust REJECT 
Adjust the cut-off level for the low velocity of TVI to be 
discarded when generating the color image.
12. Adjust FREQUENCY 
Enables the adjustment of the transmission frequency 