GE HEALTHCARE
DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL
Chapter 7 - Diagnostics/Troubleshooting 7 - 27
7-6-5-1 GSUI Acquisition Diagnostics - Test All
One test is available, Test All.
NOTE: Disconnect all probes before running this test.
Test All runs a sequential test as listed below.
NOTE: Relay Board (RLB) and Chain are not tested in this test. Some parts of the Transmitter (TX)
Board are tested during the FEC Board test, but other parts of the TX Board aren’t tested at all.
A.) Units with FEP1:
1.) PC2IP
2.) IMP (Image Port Board)
3.) VME (VME data bus on the Back Plane)
4.) SDP (Spectrum Doppler Processor)
5.) RFT (Radio Frequency & Tissue Board)
6.) FEC (Front-End Controller Board) (Some parts of TX board is also tested)
7.) BF (Beamformer Boards)
8.) RX (Receiver Board)
9.) TXP (TX Power)
B.) Units with FEP2:
1.) PC2IP
2.) RFI (Radio Frequency Interface Board)
3.) BF (Beamformer Boards)
4.) RX (Receiver Board)
5.) TXP (TX Power)