GE HEALTHCARE
DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL
Chapter 7 - Diagnostics/Troubleshooting 7 - 29
7-6-5-4 Control Frame
Contains the user interface elements used for:
Diagnostic control
Operator feedback
a.) Execute Button
This button has two modes - each with appropriate text:
* Execute - to start the diagnostic, and
* Abort - to stop a diagnostic
b.) Loop Count
This is an editable text field that will only accept numeric values with 4 digits or less. When the
button described in Execute Button step a (above) is configured as an “execute” button and
pressed, the loop count field will be queried to determine the number of times to execute the
diagnostic.
c.) Progress Indicator
Displays a graphical progress indicator for the user.
d.) Short Text Message
Displays either a starting message or aborting message, as well as the diagnostic completion
status.
e.) Control Frame Background Color
Initially gray, the Control Frame background color changes upon completion of a diagnostic
to indicate completion status.
* Fail = Red