DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL
7 - 70 Section 7-8 - Acquisition Diagnostics
7-8-3-5 Test Loops, System (cont’d)
7-8-3-6 Mode (Calibration/Test Screen Selection)
Table 7-7 Test Loops, System - Test All, System - Break (FEP2)
FIELD KEY DESCRIPTION
TEST
LOOPS
-
IN SOME CASES YOU MAY WANT TO DO THE SAME TEST SEVERAL TIMES.
ENTER A NUMBER TO INDICATE HOW MANY TIMES YOU WANT THE TEST TO REPEAT ITSELF.
DEFAULT VALUE = 1..
SYSTEM TEST ALL
USE TEST ALL
TO START A SEQUENCE OF ALL THE TESTS AVAILABLE IN THE TEST SCREEN.
THE DIAGNOSTIC TEST IS DONE IN THIS SEQUENCE ON UNITS WITH FEP2;
1.) PC2IP
2.) RFI (RADIO INTERFACE BOARD)
3.) BF (BEAM FORMER BOARDS)
4.) RX (RECEIVER BOARD)
5.) TXP (TX POWER)
NOTE: Relay Board (RLB) and Chain are not tested in this test.
NOTE: The Transmitter Board (TX) don’t have its own test, but some parts of
the TX board is tested during the FEC test.
NOTE: If the system must be rebooted, do so. Continue with the testing from
the point at which you were required to stop.
SYSTEM BREAK
IF YOU HAVE STARTED A LOOPED TEST, YOU MAY SELECT BREAK
TO STOP THE TEST.
THE TEST WILL CONTINUE UNTIL THE CURRENT TEST HAS FINISHED.
Table 7-8 Mode (Calibration/Test Screen selection)
FIELD KEY DESCRIPTION
MODE
The buttons in this field are used to select the Calibration Screen or the Test Screen.
Calibration
Select this button to select the Calibration Screen.
Test
Select this button to select the Test Screen.