DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL
Chapter 7 - Diagnostics/Troubleshooting 7 - 69
7-8-3-4 System Test Screen Description
7-8-3-5 Test Loops, System
Figure 7-74 System Test Screen, Upper Part
Table 7-6 Test Loops, System - Test All, System - Break (FEP1)
FIELD KEY DESCRIPTION
TEST
LOOPS
-
IN SOME CASES YOU MAY WANT TO DO THE SAME TEST SEVERAL TIMES.
ENTER A NUMBER TO INDICATE HOW MANY TIMES YOU WANT THE TEST TO REPEAT ITSELF.
DEFAULT VALUE = 1..
SYSTEM TEST ALL
USE TEST ALL
TO START A SEQUENCE OF ALL THE TESTS AVAILABLE IN THE TEST SCREEN.
THE DIAGNOSTIC TEST IS DONE IN THIS SEQUENCE ON UNITS WITH FEP1;
1.PC2IP
2.IMP (IMAGE PORT BOARD)
3.VME (VME DATA BUS ON THE BACK PLANE)
4.SDP (SPECTRUM DOPPLER PROCESSOR)
5.RFT (RADIO FREQUENCY & TISSUE BOARD)
6.FEC (FRONT-END CONTROLLER BOARD)
7.BF (BEAMFORMER BOARDS)
8.RX (RECEIVER BOARD)
9.TXP (TX POWER)
NOTE: Relay Board (RLB) and Chain are not tested in this test.
NOTE: The Transmitter Board (TX) don’t have its own test, but some parts of
the TX board is tested during the FEC test.
NOTE: If the system must be rebooted, do so. Continue with the testing from
the point at which you were required to stop.
SYSTEM BREAK
IF YOU HAVE STARTED A LOOPED TEST, YOU MAY SELECT BREAK
TO STOP THE TEST.
THE TEST WILL CONTINUE UNTIL THE CURRENT TEST HAS FINISHED.
System
Test Loops
Mode