GE HEALTHCARE
DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL
8 - 26 Section 8-7 - Software Loading Procedure - Software Version v7.x, v6.x, v5.x and v4.x
8-7-12-3 Loading System Software v3.0.x on Vivid 7 with BEP4.3 (cont’d)
4.) In this step, you can choose if you want to do a complete Vivid 7 software installation that overwrites
all data, or if you only want to update or reinstall the software on the C: partition (C: drive).
Usually it should be OK to select “B” to update the C: drive only. This will keep Patient Archive and
Presets intact. Continue with either step a (if possible) or step b (if you have to):
NOTE: If you select “A” below, all program data, Patient Archive and Presets will be erased.
a.) ONLY select “A” if a “B” update cannot be used: Select A to do a complete Vivid 7 software
installation.
The reload process start automatically. When done, a message tells you to initiate Application
software installation. Continue with step 5.
b.) Select “B” to update/re-install the software on the C: partition. (Only data on the C: partition will
be erased. This will keep Patient Archive and Presets intact.)
The reload process starts automatically. When done, a message tells you to initiate Application
software installation.
5.) Press Y to confirm.
The Vivid 7 reboots and the Application software installation starts automatically. When completed, the
Vivid 7 reboots normally.
If you selected A above, the User Presets, the System Configurations and the local Patient Archive
has been reset to the factory default settings.
If you selected B above, the User Presets, the System Configurations and the local Patient Archive
remains as before the software reload.
Figure 8-22 System Recovery Menu
NOTICE
If you select “A” in the next step, ALL existing software and data will be erased. If backup has
not been performed as described earlier in this manual; ALL DATA like Patient Database,
System Configuration and User Configurations (Customer Presets) will be lost.
System Recovery Menu
A. Full reload (erase all partitions)
B. Reload Partition C:\ only (Keep Archive)
C. Reload Application Software
R. Restart System
Select (A, B, C,,R) _