DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL
8 - 42 Section 8-8 - Software Loading Procedure - Software Version v3.x
8-8-6-2 Backup of Patient Archive and System Configurations to MO Disks (cont’d)
7.) Select Backup Now.
a.) The system will first check if the MO disk is inserted. If not, a dialogue window is displayed
prompting you to insert an MO disk. Select OK
to resume backup.
b.) The system will then check if the MO disk is formatted. If not, a dialogue window is displayed
prompting you to format the disk.
A progress dialogue is displayed showing the current operation being performed.
If the amount of data exceeds the capacity of the MO disk, the user is prompted to insert a new
(formatted) MO disk. At the end of the process, a status for each item is displayed in the Result column.
If errors occur during backup, a list of the patient records/ examinations that could not be backed up is
displayed.